  4. Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis.
 
  • Détails
Titre

Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis.

Type
article
Institution
UNIL/CHUV/Unisanté + institutions partenaires
Périodique
Physics in Medicine & Biology  
Auteur(s)
Monnin, P.
Auteure/Auteur
Marshall, N.W.
Auteure/Auteur
Bosmans, H.
Auteure/Auteur
Bochud, F.O.
Auteure/Auteur
Verdun, F.R.
Auteure/Auteur
Liens vers les personnes
Bochud, François  
Verdun, Francis  
Monnin, Pascal  
Liens vers les unités
Radiophysique  
ISSN
1361-6560
Statut éditorial
Publié
Date de publication
2011
Volume
56
Numéro
14
Première page
4221
Dernière page/numéro d’article
4238
Langue
anglais
Résumé
Assessment of image quality for digital x-ray mammography systems used in European screening programs relies mainly on contrast-detail CDMAM phantom scoring and requires the acquisition and analysis of many images in order to reduce variability in threshold detectability. Part II of this study proposes an alternative method based on the detectability index (d') calculated for a non-prewhitened model observer with an eye filter (NPWE). The detectability index was calculated from the normalized noise power spectrum and image contrast, both measured from an image of a 5 cm poly(methyl methacrylate) phantom containing a 0.2 mm thick aluminium square, and the pre-sampling modulation transfer function. This was performed as a function of air kerma at the detector for 11 different digital mammography systems. These calculated d' values were compared against threshold gold thickness (T) results measured with the CDMAM test object and against derived theoretical relationships. A simple relationship was found between T and d', as a function of detector air kerma; a linear relationship was found between d' and contrast-to-noise ratio. The values of threshold thickness used to specify acceptable performance in the European Guidelines for 0.10 and 0.25 mm diameter discs were equivalent to threshold calculated detectability indices of 1.05 and 6.30, respectively. The NPWE method is a validated alternative to CDMAM scoring for use in the image quality specification, quality control and optimization of digital x-ray systems for screening mammography.
PID Serval
serval:BIB_E78C7646DCFD
DOI
10.1088/0031-9155/56/14/003
PMID
21701050
WOS
000292303100004
Permalien
https://iris.unil.ch/handle/iris/256846
Date de création
2011-07-15T09:13:30.483Z
Date de création dans IRIS
2025-05-21T07:07:32Z
Fichier(s)
En cours de chargement...
Vignette d'image
Nom

Monnin_PhysMedBiol_2011_PartII.pdf

Version du manuscrit

published

Licence

https://creativecommons.org/licenses/by-nd/4.0

Taille

1.59 MB

Format

Adobe PDF

PID Serval

serval:BIB_E78C7646DCFD.P001

URN

urn:nbn:ch:serval-BIB_E78C7646DCFD0

Somme de contrôle

(MD5):559afa5355fab17312f99d33e280717a

  • Copyright © 2024 UNIL
  • Informations légales