  4. Evaluation of the Sysmex XQ-320 three-part differential haematology analyser and its flagging capabilities.
 
  • Détails
Titre

Evaluation of the Sysmex XQ-320 three-part differential haematology analyser and its flagging capabilities.

Type
article
Institution
UNIL/CHUV/Unisanté + institutions partenaires
Périodique
Journal of Clinical Laboratory Analysis  
Auteur(s)
Coutaz, C.
Auteure/Auteur
Mamin, A.
Auteure/Auteur
Mintzas, K.
Auteure/Auteur
Liens vers les unités
Laboratoires d'hématologie  
ISSN
1098-2825
Statut éditorial
Publié
Date de publication
2024-02
Volume
38
Numéro
4
Première page
e25017
Peer-reviewed
Oui
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
Three-part differential (3PD) haematology analysers offer a quick, easy-to-use and economical way to acquire important information about a patient's physiology. In this study, we evaluated a new 3PD analyser, the Sysmex XQ-320, investigated its comparability with its predecessor (Sysmex XP-300) and the five-part differential analyser Sysmex XN-9000, and explored its flagging potential.
Analytical performance studies were conducted for repeatability, within-laboratory precision, between-day precision, carry-over and linearity with fresh blood and QC material. Method comparison was performed in 493 samples comparing XQ-320 with XP-300, using the XN-9000 as the gold standard.
The XQ-320 excelled manufacturer's specifications in the analytical performance studies, except for MXD in within-laboratory and between-day precisions using the QC material level 1. The XQ-320 showed correlation values greater than 0.94 with XN-9000 for the majority of the 20 reportable parameters (MXD# 0.891, MXD% 0.898 and MCHC 0.849). Improvements over the XP-300 were observed in WBC in the leucocytopenic range (bias -0.038 vs. -0.097) and PLT (bias 2.568 vs. -7.877, intercept 3.880 vs. -8.845). Concordance between XQ-320 and XP-300 was 91.9% for the WBC histogram abnormal distribution flag and 95.3% for the PLT flag. Patterns of increased neutrophils and decreased mixed cells on the XQ-320 were observed in samples that raised a flag on XN-9000.
The XQ-320 showed excellent analytical performance, and very good to excellent correlation with XN-9000 with improvements over XP-300. Flagging combined with parameter patterns identified additional suspected abnormal samples, thus making the XQ-320 an excellent solution for laboratories utilising 3PD analysers.
Sujets

Humans

Hematology

Laboratories

Nonoxynol

Blood Cell Count/meth...

Reproducibility of Re...

analytical performanc...

correlation

flagging

haematology analyser

three-part differenti...

PID Serval
serval:BIB_31CC15F1E63D
DOI
10.1002/jcla.25017
PMID
38396348
WOS
001169873600001
Permalien
https://iris.unil.ch/handle/iris/60132
Open Access
Oui
Date de création
2024-03-01T12:11:19.476Z
Date de création dans IRIS
2025-05-20T15:32:00Z
Fichier(s)
En cours de chargement...
Vignette d'image
Nom

38396348_BIB_31CC15F1E63D.pdf

Version du manuscrit

published

Licence

https://creativecommons.org/licenses/by/4.0

Taille

1.24 MB

Format

Adobe PDF

PID Serval

serval:BIB_31CC15F1E63D.P001

URN

urn:nbn:ch:serval-BIB_31CC15F1E63D7

Somme de contrôle

(MD5):5b4835bfbcf2db9b8d14d4263de49700

  • Copyright © 2024 UNIL
  • Informations légales