  4. A new technique to measure micromotion distribution around a cementless femoral stem.
 
  • Détails
Titre

A new technique to measure micromotion distribution around a cementless femoral stem.

Type
article
Institution
UNIL/CHUV/Unisanté + institutions partenaires
Périodique
Journal of Biomechanics  
Auteur(s)
Gortchacow, M.
Auteure/Auteur
Wettstein, M.
Auteure/Auteur
Pioletti, D.P.
Auteure/Auteur
Terrier, A.
Auteure/Auteur
Liens vers les personnes
Wettstein, Michael  
Pioletti, Dominique  
Liens vers les unités
Orthopédie et traumatologie  
ISSN
1873-2380[electronic], 0021-9290[linking]
Statut éditorial
Publié
Date de publication
2011
Volume
44
Numéro
3
Première page
557
Dernière page/numéro d’article
560
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Résumé
The interfacial micromotion is closely associated to the long-term success of cementless hip prostheses. Various techniques have been proposed to measure them, but only a few number of points over the stem surface can be measured simultaneously. In this paper, we propose a new technique based on micro-Computer Tomography (μCT) to measure locally the relative interfacial micromotions between the metallic stem and the surrounding femoral bone. Tantalum beads were stuck at the stem surface and spread at the endosteal surface. Relative micromotions between the stem and the endosteal bone surfaces were measured at different loading amplitudes. The estimated error was 10μm and the maximal micromotion was 60μm, in the loading direction, at 1400N. This pilot study provided a local measurement of the micromotions in the 3 direction and at 8 locations on the stem surface simultaneously. This technique could be easily extended to higher loads and a much larger number of points, covering the entire stem surface and providing a quasi-continuous distribution of the 3D interfacial micromotions around the stem. The new measurement method would be very useful to compare the induced micromotions of different stem designs and to optimize the primary stability of cementless total hip arthroplasty.
PID Serval
serval:BIB_5CC5B07BA0D9
DOI
10.1016/j.jbiomech.2010.09.023
PMID
20934705
WOS
000287551000031
Permalien
https://iris.unil.ch/handle/iris/38142
Date de création
2011-02-02T12:18:50.896Z
Date de création dans IRIS
2025-05-20T13:44:16Z