  4. Distal locking of femoral nails. Mathematical analysis of the appropriate targeting range.
 
  • Détails
Titre

Distal locking of femoral nails. Mathematical analysis of the appropriate targeting range.

Type
article
Institution
Externe
Périodique
Orthopaedics & Traumatology: Surgery & Research  
Auteur(s)
Moor, B.K.
Auteure/Auteur
Ehlinger, M.
Auteure/Auteur
Arlettaz, Y.
Auteure/Auteur
Liens vers les personnes
Arlettaz, Yvan  
ISSN
1877-0568
Statut éditorial
Publié
Date de publication
2012
Volume
98
Numéro
1
Première page
85
Dernière page/numéro d’article
89
Langue
anglais
Notes
Publication types: Comparative Study ; Journal Article Publication Status: ppublish
Résumé
BACKGROUND: During the last decades, intramedullary nailing has become the standard treatment for diaphyseal fractures of long bones. Numerous innovative techniques and devices have been proposed to simplify distal locking. Each has its own limitations and, as a result, the fluoroscopy-dependent "free-hand technique" remains the most popular method. However, radiation exposure to the patient and operating room staff remains a concern.
METHODS: Before the development of a new radiation-independent, nail-mounted targeting system, we mathematically analyzed the aiming accuracy that such a system has to achieve. The correctness of this mathematical model was evaluated using a mechanical testing apparatus.
FINDINGS: We found a quite large targeting range for the unimpeded passage of the drill bit through the locking hole of a given nail. Important degrees of nail bending can thereby be compensated. As predicted by the mathematical formula, a 4-mm drill bit passed the distal locking hole of a 320/11mm femoral nail up to a deflection of ±13mm in the coronal plane.
INTERPRETATION: This mathematical model can be considered to be an additional tool for the development of new targeting devices. Combining our mathematical model with data previously published, not only torsional deformation along the longitudinal axis of the nail but also bending in the coronal plane can approximately be neglected. Hence, the three-dimensional aiming process can be simplified to the determination of the interlocking hole of the nail in the sagittal plane provided that the insertion-induced nail deformation in vivo stays in the range of that observed in vitro.
LEVEL OF EVIDENCE: Level III. Basic sciences control study.
Sujets

Biomechanical Phenome...

Bone Nails

Bone Screws

Cadaver

Equipment Design

Femoral Fractures/phy...

Femoral Fractures/sur...

Fracture Fixation, In...

Humans

Imaging, Three-Dimens...

Materials Testing/ins...

Models, Theoretical

Prosthesis Design

Reproducibility of Re...

PID Serval
serval:BIB_CB3FC885515B
DOI
10.1016/j.otsr.2011.08.010
PMID
22264565
WOS
000300392900012
Permalien
https://iris.unil.ch/handle/iris/217815
Open Access
Oui
Date de création
2014-09-23T17:40:42.807Z
Date de création dans IRIS
2025-05-21T04:02:07Z
Fichier(s)
En cours de chargement...
Vignette d'image
Nom

BIB_CB3FC885515B.P001.pdf

Version du manuscrit

published

Taille

552.16 KB

Format

Adobe PDF

PID Serval

serval:BIB_CB3FC885515B.P001

Somme de contrôle

(MD5):c8adaef7c4dbddc70ce244049d7d6b43

  • Copyright © 2024 UNIL
  • Informations légales