  4. Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement.
 
  • Détails
Titre

Computer-assisted cup placement techniques in total hip arthroplasty improve accuracy of placement.

Type
article
Institution
UNIL/CHUV/Unisanté + institutions partenaires
Périodique
Clinical Orthopaedics and Related Research  
Auteur(s)
Jolles, B.M.
Auteure/Auteur
Genoud, P.
Auteure/Auteur
Hoffmeyer, P.
Auteure/Auteur
Liens vers les personnes
Jolles, Brigitte  
Liens vers les unités
Orthopédie et traumatologie  
ISSN
0009-921X
Statut éditorial
Publié
Date de publication
2004
Numéro
426
Première page
174
Dernière page/numéro d’article
179
Peer-reviewed
Oui
Langue
anglais
Résumé
Malposition of the acetabular component during hip arthroplasty increases the occurrence of impingement, reduces range of motion, and increases the risk of dislocation and long-term wear. To prevent malpositioned hip implants, an increasing number of computer-assisted orthopaedic systems have been described, but their accuracy is not well established. The purpose of this study was to determine the reproducibility and accuracy of conventional versus computer-assisted techniques for positioning the acetabular component in total hip arthroplasty. Using a lateral approach, 150 cups were placed by 10 surgeons in 10 identical plastic pelvis models (freehand, with a mechanical guide, using computer assistance). Conditions for cup implantations were made to mimic the operating room situation. Preoperative planning was done from a computed tomography scan. The accuracy of cup abduction and anteversion was assessed with an electromagnetic system. Freehand placement revealed a mean accuracy of cup anteversion and abduction of 10 degrees and 3.5 degrees, respectively (maximum error, 35 degrees). With the cup positioner, these angles measured 8 degrees and 4 degrees (maximum error, 29.8 degrees), respectively, and using computer assistance, 1.5 degrees and 2.5 degrees degrees (maximum error, 8 degrees), respectively. Computer-assisted cup placement was an accurate and reproducible technique for total hip arthroplasty. It was more accurate than traditional methods of cup positioning.
Sujets

Arthroplasty, Replace...

Electromagnetic Pheno...

Humans

Models, Anatomic

Reproducibility of Re...

Surgery, Computer-Ass...

PID Serval
serval:BIB_43C3FBF9DD29
DOI
10.1097/01.blo.0000141903.08075.83
PMID
15346070
WOS
000223708600028