Titre
Leadless pacing using the transcatheter pacing system (Micra TPS) in the real world: initial Swiss experience from the Romandie region.
Type
article
Institution
UNIL/CHUV/Unisanté + institutions partenaires
Périodique
Auteur(s)
Valiton, V.
Auteure/Auteur
Graf, D.
Auteure/Auteur
Pruvot, E.
Auteure/Auteur
Carroz, P.
Auteure/Auteur
Fromer, M.
Auteure/Auteur
Bisch, L.
Auteure/Auteur
Tran, V.N.
Auteure/Auteur
Cook, S.
Auteure/Auteur
Scharf, C.
Auteure/Auteur
Burri, H.
Auteure/Auteur
Liens vers les personnes
Liens vers les unités
ISSN
1532-2092
Statut éditorial
Publié
Date de publication
2019-02-01
Volume
21
Numéro
2
Première page
275
Dernière page/numéro d’article
280
Peer-reviewed
Oui
Langue
anglais
Notes
Publication types: Journal Article
Publication Status: ppublish
Publication Status: ppublish
Résumé
Leadless pacemakers are implanted in Switzerland since June 2015. Large worldwide registries have shown high implant success, low complication rates, and good electrical parameters up to 12 months' follow-up. However, data are scarce outside the investigational setting. The purpose of this study is to assess the real-world experience regarding clinical safety and efficacy of Micra TPS (transcatheter pacing system) leadless pacemakers.
Retrospective observational, multi-centre study designed to assess initial safety and efficacy of the Micra TPS in the Swiss Romande region. A total of 92 patients were included from four different centres with an implantation success rate of 97.8% (90 of 92). Thresholds were overall low at implantation (median 0.38 V/0.24 ms, ranging from 0.13 to 2.88 V/0.24 ms) and remained stable over 1-year follow-up. The perioperative serious adverse event rate was 6.5% in six patients which lead to prolonged hospitalization in five patients and death in one patient. In addition, three further major events (3.3%) occurred during an average follow-up of 1 year, requiring implantation of a standard transvenous pacemaker in two patients, and surgical explantation of the Micra TPS in one patient due to intractable ventricular tachycardia.
Leadless pacemakers are a valuable adjunct for treating selected patients requiring single-chamber pacing. However, in this initial experience, major complication rates were high (9.8%). The implant procedure requires proper training and should be performed in an adequate setting.
Retrospective observational, multi-centre study designed to assess initial safety and efficacy of the Micra TPS in the Swiss Romande region. A total of 92 patients were included from four different centres with an implantation success rate of 97.8% (90 of 92). Thresholds were overall low at implantation (median 0.38 V/0.24 ms, ranging from 0.13 to 2.88 V/0.24 ms) and remained stable over 1-year follow-up. The perioperative serious adverse event rate was 6.5% in six patients which lead to prolonged hospitalization in five patients and death in one patient. In addition, three further major events (3.3%) occurred during an average follow-up of 1 year, requiring implantation of a standard transvenous pacemaker in two patients, and surgical explantation of the Micra TPS in one patient due to intractable ventricular tachycardia.
Leadless pacemakers are a valuable adjunct for treating selected patients requiring single-chamber pacing. However, in this initial experience, major complication rates were high (9.8%). The implant procedure requires proper training and should be performed in an adequate setting.