  4. A Monte Carlo-based procedure for independent monitor unit calculation in IMRT treatment plans.
 
  • Détails
Titre

A Monte Carlo-based procedure for independent monitor unit calculation in IMRT treatment plans.

Type
article
Institution
UNIL/CHUV/Unisanté + institutions partenaires
Périodique
Physics in Medicine & Biology  
Auteur(s)
Pisaturo, O.
Auteure/Auteur
Moeckli, R.
Auteure/Auteur
Mirimanoff, R.O.
Auteure/Auteur
Bochud, F.O.
Auteure/Auteur
Liens vers les personnes
Mirimanoff, René-Olivier  
Pisaturo, Olivier  
Bochud, François  
Moeckli, Raphaël  
Liens vers les unités
Radiophysique  
Radio-oncologie  
ISSN
0031-9155
Statut éditorial
Publié
Date de publication
2009
Volume
54
Numéro
13
Première page
4299
Dernière page/numéro d’article
4310
Peer-reviewed
Oui
Langue
anglais
Résumé
Intensity-modulated radiotherapy (IMRT) treatment plan verification by comparison with measured data requires having access to the linear accelerator and is time consuming. In this paper, we propose a method for monitor unit (MU) calculation and plan comparison for step and shoot IMRT based on the Monte Carlo code EGSnrc/BEAMnrc. The beamlets of an IMRT treatment plan are individually simulated using Monte Carlo and converted into absorbed dose to water per MU. The dose of the whole treatment can be expressed through a linear matrix equation of the MU and dose per MU of every beamlet. Due to the positivity of the absorbed dose and MU values, this equation is solved for the MU values using a non-negative least-squares fit optimization algorithm (NNLS). The Monte Carlo plan is formed by multiplying the Monte Carlo absorbed dose to water per MU with the Monte Carlo/NNLS MU. Several treatment plan localizations calculated with a commercial treatment planning system (TPS) are compared with the proposed method for validation. The Monte Carlo/NNLS MUs are close to the ones calculated by the TPS and lead to a treatment dose distribution which is clinically equivalent to the one calculated by the TPS. This procedure can be used as an IMRT QA and further development could allow this technique to be used for other radiotherapy techniques like tomotherapy or volumetric modulated arc therapy.
Sujets

Algorithms

Computer Simulation

Humans

Models, Biological

Monte Carlo Method

Radiometry/methods

Radiotherapy Dosage

Radiotherapy Planning...

Radiotherapy, Conform...

Reproducibility of Re...

Sensitivity and Speci...

PID Serval
serval:BIB_B973549F0199
DOI
10.1088/0031-9155/54/13/022
PMID
19531844
WOS
000267137200022
Permalien
https://iris.unil.ch/handle/iris/178539
Date de création
2009-09-29T08:22:34.288Z
Date de création dans IRIS
2025-05-21T00:46:34Z
  • Copyright © 2024 UNIL